The Protein Clear HR LabChip Touch chip is used with the LabChip® GXII Touch Protein Clear™ HR assay on the LabChip® GXII Touch™ instrument. The LabChip GXII Touch Protein Clear HR assay is designed for the high-resolution protein purity assessment of monoclonal antibodies (mAb) within research and development, enabling the detection and visualization of impurities within a sample, down to 5 ng/μL.
For research use only. Not for use in diagnostic procedures.
Feature | Specification |
---|---|
Application | Protein Analysis |
Unit Size | 1 chip |
Feature | Specification |
---|---|
Application | Protein Analysis |
Unit Size | 1 chip |
Protein Impurity Analysis in seconds
The Protein Clear HR LabChip Touch HT runs 96 or fewer sample batch sizes per prime on the LabChip GXII Touch HT instrument. The Protein Clear HR assay is designed for the high-resolution protein purity assessment of monoclonal antibodies (mAb) within research and development, enabling the detection and visualization of impurities within a sample, down to 5 ng/μL. It has a total sample count of 400, in 8 primes. With the analytical power and speed of the LabChip® GXII Touch system and IntelliChip™ assay optimization technology, you can accelerate your protein development workflows with unmatched throughput and reproducibility.
Application |
Protein Analysis
|
---|---|
Automation Compatible |
Yes
|
Instrument Compatibility |
LabChip GXII Touch HT
|
Sample Type |
Protein
|
Shipping Conditions |
Shipped in Blue Ice Soft
|
Technology |
Microfluidic Electrophoresis
|
Unit Size |
1 chip
|
Parameter | Specification |
---|---|
Sizing range | 14-250 kDa |
Maximum sample concentration | 2000 ng/μL |
Sizing resolution* | Resolution > 1.0 for VeriMAb reference standard |
Sizing precision RSD (CV) | <2% |
Relative migration time precision RSD (CV) | <2% |
Percent purity reproducibility | <0.5% Main non-reduced IgG <5% all other peaks |
Sensitivity (LOD) | 5 ng/μL (mAb, non-reduced main peak) |
Linear concentration range | 10 – 1000 ng/ μL (mAb, non-reduced main peak) |
Linearity (R^2) | >0.995 |
Separation time per sample | 65 seconds |
Maximum number of samples per calibration | 96 |
Chip lifetime | 400 samples |
Reagent kit primes | 10 |
Minimum sample volume | 2 μL |
Are you looking for resources, click on the resource type to explore further.
In this application note mAbs were evaluated using both the Protein Clear™ HR assay and an orthogonal method, the SCIEX® PA800...
Read the LabChip® GXII Touch™ protein characterization system product note to learn about an automated alternative to traditional...
Capillary electrophoresis (CE), more specifically CE-SDS, is becoming an increasingly popular and routinely used method for...
Are you looking for technical documents related to the product? We have categorized them in dedicated sections below. Explore now.
We are here to answer your questions.