Overview
Assays involving radiochemicals can be run in various medium- and high-throughput plate-based formats, including filtration assays (commonly used for thymidine incorporation and radioligand binding assays), liquid scintillation counting assays, scintillant coated-plate assays (commonly used for HPLC fraction analysis or 51Cr release assays), and proximity assays (using SPA beads or Cytostar-T™ plates). Plate recommendations for each assay format are detailed in the sections below. Radiometric assays that measure signal coming from a beta-emitting radioisoe (such as 3H, 14C, 32P, and 35S) or a radioisoe that releases Auger electrons (such as 125I and 51Cr) typically rely on addition of a scintillant, which captures the energy from the emitted beta particle and, in turn, releases energy in the form of photons that can be measured on radiometric instruments (such as the Revvity MicroBeta or Count Plate Counters) that use a photomultiplier tube for detection. Because of this, plate selection for these assays will factor in some of the same considerations as for standard luminescence assays. Liquid scintillation counting assays require the use of liquid scintillation cocktails, which can sometimes contain harsh organic solvents. For these assays, compatibility of the plate material with the cocktail solvent should be a consideration. Additional considerations such as whether the detection instrument is reading from the and/or bottom of the plate should also be factored in when choosing a plate. Clear-bottom plates allow for bottom-reading and coincidence measurements on instruments such as the MicroBeta plate counter. Clear-bottom plates with black or white wells can be used in conjunction with BackSeal™ Adhesive Bottom Seal to reduce cross-talk in -reading measurements. Solid opaque plates need to be measured with -reading instruments, such as the Count system (the MicroBeta counter can also support -reading measurements).
Plate color
Signal
The use of white plates will result in higher signal for photon measurement assays (for example, radiometric assays that use a scintillant to convert energy from a beta-emitting or Auger electron-emitting radioisoe into photons), whereas black plates can "quench" the signal by absorbing some of the light produced by the assay. For most radiometric applications, we recommend the use of white plates, or white-walled plates with clear bottoms for bottom-reading instruments. However, if you are working with an assay that gives a strong signal, black or gray plates may be helpful in reducing background from phosphorescence (see next section).
Phosphorescence and dark adaptation
For assays that involve use of a scintillation cocktail or other scintillant (SPA bead, Cytostar-T plate, LumaPlate™, MeltiLex™), phosphorescence may be a potential issue. Radiometric assays that measure signal coming from a beta-emitting radioisoe (such as 3H, 14C, 32P, and 35S) typically rely on addition of a scintillant, which captures the energy from the emitted beta particle and, in turn, releases energy in the form of photons that can be measured on radiometric instruments that use a photomultiplier tube for detection. Assays involving radioisoes that emit Auger electrons (such as 125I and 51Cr) can also use scintillants in a similar manner.
Background phosphorescence is a consideration when choosing between white or black plates for light-producing (photon-producing) assays. Phosphorescence is the emission of light by a substance, resulting from stored energy. Certain components in a buffer or sample may phosphoresce, and plastic from the microplate itself can also phosphoresce. Phosphorescence can lead to increased background, which could potentially have a negative impact on a given assay. Black or gray microplates intrinsically exhibit less phosphorescence than white microplates, and may be desirable for particular assays. Alternatively, assays run in white microplates can be "dark-adapted" by shielding the microplate from light prior to reading the plate in order to reduce background phosphorescence. You can start by dark-adapting the plate for 5 minutes. The amount of time required for dark adaptation will vary from assay-to-assay, and should be optimized.
Cross-talk
There are two types of cross-talk: optical and isoic (or radiometric) cross-talk. Optical cross-talk occurs when light from one well travels to another well where it is detected and adds non-specific counts to that well. Isoic cross-talk occurs when high energy beta particles from one well travel to adjacent wells where they interact with the scintillant in that well, adding to non-specific counts. This can be significant when high energy isoes are used, such as 32P.
Microplates recommended for use in radiometric assays are thick-walled rigid plates containing colorants which produce an opaque white, black, or gray finish. These pigments, in conjunction with wall thickness, help attenuate the transmission of radioactive decay energy between neighboring wells, which would otherwise lead to "cross-talk". The use of opaque (solid-color) microplates can also reduce optical cross-talk to negligible levels.
Microplates for filtration assays
We offer glass fiber UniFilter™ plates (GF/B and GF/C) for filtration assays. The difference between GF/B and GF/C filters is the filter thickness (GF/B is 680 µm and GF/C is 260 µm), resulting in a different pore size (1 µm or 1.2 µm, respectively). The size of the pore determines how much sample is retained on the filter, and how much is filtered as a waste material. Assays involving cell membrane preparations can be run using either GF/B or GF/C filters. For whole cell-based assays, we recommend GF/C filters to prevent clogging.
UniFilter plates are also available with a polyethyleneimine (PEI) coating. PEI is a cationic polymer that can neutralize the negative charge of the glass fiber filter, and can be used when a positively-charged ligand or other reagent exhibits high non-specific binding to the glass fiber filter.
UniFilter plates should be used in conjunction with BackSeal (catalog number 6005199 for white, catalog number 6005189 for black) and Seal™-A adhesive seals. BackSeal adhesive seal fits onto the bottom of the filter plate, which should be applied before addition of scintillation cocktail. Seal-A adhesive seal fits on of the plate to prevent spills/contamination when placing the plates into the high-throughput detection instrument.
Both the Revvity Count and MicroBeta plate counters can be used to measure filtration assays.
Table 1: Filter plate products
Plate type | Filter type | Well format | Application | Plates | Catalog number |
---|---|---|---|---|---|
UniFilter plate | GF/C | 96-well | Labeled cell assays (Max, loading typically 250,000 cells) | 50 | 6055690 |
PEI-coated UniFilter plate | GF/C | 96-well | Labeled cell assays (Max. loading typically 250,000 cells) | 50 | 6055090 |
*The Harvest Plate is a 96-well filter plate designed for use with the Tomtec™ Mach IIIW automatic harvester. The Harvest Plate includes a retaining plate that is designed to isolate each of the 96 filter areas and also to support the filter material. The Harvest Plate is made up with GF/C glass fiber. Other filter types are available on request. The external dimensions of the Harvest Plate are identical to a standard 96-well microplate and it is therefore highly suitable for use in automated systems.
Alternatively, we offer filtermats that can be placed into cassettes compatible with the high-throughput detection instrument being used (Table 2) or filtermats that are used with the FilterMate Universal Harvester or in the OmniFilter™ Plate Assembly (Table 3).
Table 2. Filtermat products (to be used with cassette to adapt to high-throughput detector)
Filtermat type | Well format | Application | Filters | Catalog Number |
---|---|---|---|---|
GF/B | 24 | Receptor-ligand binding assays (Max. loading 400 µg) | 50 | 50 |
GF/B | 96 | Receptor-ligand binding assays (Max. loading 200 µg) | 50 | 1450-521 |
GF/C | 96 | Labeled cell assays (Max. loading typically 250,000 cells) | 100 | 1450-421 |
Nylon | 96 | RNA/DNA blots (0.45 µm membrane) | 25 | 1450-423 |
Table 3. Filtermat products to be used with FilterMate Harvester or OmniFilter plate to adapt to high-throughput detector
Filtermat name | Material/Application | Filters | Catalog Number |
---|---|---|---|
Easytab-C filtermat | Receptor-ligand binding assays or labeled cell assays (GF/C filter); has polyester backing material1 and self-aligning removable tabs for ease of mounting filter in OmniPlate | 100 | 6005422 |
Standard self-aligning filtermat | Receptor-ligand binding assays or labeled cell assays; Glass fiber filter; particle retention of 1.5 µm; Has no polyester backing material1 | 100 | 6005416 |
Microplates for liquid scintillation counting assays
Liquid scintillation counting assays require the use of liquid scintillation cocktails, which can sometimes contain harsh organic solvents. For these assays, compatibility of the plate material with the cocktail solvent should be a consideration. PicoPlates are composed of solvent-resistant Barex® plastic. OptiPlates, Isoplates, and VisiPlates are composed of polystyrene, and are less suitable for use with liquid scintillation cocktails that contain harsher solvents. The Count and MicroBeta instruments from Revvity can be used to measure liquid scintillation assays.
OptiPlate™ microplates
OptiPlate microplates are standard, highly-versatile polystyrene microplates. White or black OptiPlate microplates can be used in radiometric detection assays. OptiPlate microplates are solid (opaque) in color with no transparency, and therefore require measurement in -reading microplate readers. OptiPlate microplates are available in 24-well, 96-well, 384-well and 1536-well formats.
Isoplate™ microplates (untreated)
Clear-well Isoplate microplates are polystyrene plates that have a clear bottom, while the sides of each well are either black or white. This makes clear-well Isoplate microplates suitable for bottom-reading instruments or coincidence measurements. Isoplate microplates are manufactured by first molding 96 clear wells at a time, then molding a black or white frame around the clear wells. This makes the white- or black-colored well extend to the same depth as the clear well base, and can help reduce cross-talk in bottom-reading assays. Clear-well Isoplate microplates were developed for coincidence counting in a MicroBeta instrument (reading from and bottom coincidentally). Isoplates are only available in 96-well format.
VisiPlate™ microplates (non-TC treated)
VisiPlate microplates are polystyrene plates that have a clear-bottom base, with the sides of each well a solid white or black in color. The clear-bottom base of the plate allows for bottom-read measurements (i.e., when the plate detector is located below the plate within the plate reader). VisiPlate microplates are similar to Isoplate microplates, but are our only clear-bottom, solid-colored well plates that are offered in 24-well format. VisiPlate microplates are only offered in 24-well format.
Flexible PET plates
Flexible plates are designed for Revvity's MicroBeta liquid scintillation counter, and give the best results in liquid scintillation counting and SPA assays. The plates are made of clear polyethylene-A (PET-A) or polyethylene-G (PET-G) plastic and printed with grid lines between wells to prevent cross-talk. The plate is chemically resistant to all DIN (di-isopropyl naphthalene)-based cocktails and is suited for chromium release, Cerenkov 32P, SPA and all other general liquid scintillation applications. Plate cassettes and support frames are required to use the flexible microplates in the MicroBeta instrument. Flexible PET plates are available in 24-well and 96-well format.
LumaPlate™ microplates
The LumaPlate is a solid scintillator-coated microplate that eliminates the need for adding liquid scintillation counting cocktail. Scintillant is coated onto the bottom and sides of wells. This can offer ease of use. LumaPlate is recommended in use of HPLC or ultraPLC fraction analysis in metabolic studies and 51Cr release assays. LumaPlates are microplates that come in 96-well (shallow or deep-well) and 384-deep-well format.
Microplates for SPA assays
We recommend the use of white plates or white-walled, clear bottom plates for SPA bead-based assays.
OptiPlate™
OptiPlates are standard, highly-versatile polystyrene microplates. They are solid (opaque) in color with no transparency, and therefore require measurement in -reading microplate readers. OptiPlate microplates are available in 24-well, 96-well, 384-well and 1536-well formats.
ProxiPlate™
ProxiPlate microplates are shallow-well plates designed for low-volume assays. The bottom of the wells is raised to position the surface of the liquid in each well as close to the -reading instrument detector as possible, resulting in high signal. They are solid (opaque) in color with no transparency, and therefore require measurement in -reading microplate readers. ProxiPlate microplates are available in 96-well and 384-well formats only.
Isoplates (untreated)
Clear-well Isoplate microplates are polystyrene plates that have a clear bottom, while the sides of each well are either black or white. This makes clear-well Isoplate microplates suitable for bottom-reading instruments or coincidence measurements. Isoplates are manufactured by first molding 96 clear wells at a time, then molding a black or white frame around the clear wells. This makes the white- or black-colored well extend to the same depth as the clear well base, and can help reduce cross-talk in bottom-reading assays. Clear-well Isoplate microplates were developed for coincidence counting in a MicroBeta instrument (reading from and bottom coincidentally). Isoplate microplates are only available in 96-well format.
VisiPlate™
VisiPlate microplates are polystyrene plates that have a clear-bottom base, with the sides of each well a solid white or black in color. The clear-bottom base of the plate allows for bottom-read measurements (i.e., when the plate detector is located below the plate within the plate reader). VisiPlate microplates are similar to Isoplate microplates, but are our only clear-bottom, solid-colored well plates that are offered in 24-well format. VisiPlate microplates are only offered in 24-well format.
ScintiPlate™ and Cytostar-T plates for proximity assays
ScintiPlate, and Cytostar-T plates are coated with scintillant, and therefore do not require the addition of scintillation cocktail for your assay.
ScintiPlate™ microplates
ScintiPlate microplates are clear-bottom plates with white walls, but are otherwise similar to FlashPlate microplates, in that they are special plates that contain scintillant embedded into the plastic of the plate and are intended for use in solid-phase (coated-plate) assays. In a ScintiPlate format, some component of the assay is captured onto the bottom and sides of the well of a plate. The assay is designed so that when radiochemical binds or is utilized as a substrate, this puts the radiochemical into close proximity of the bottom and sides of the ScintiPlate microplate. The beta-energy from the radiochemical interacts with scintillant embedded in the plastic of the well, producing signal. Unbound or unused radiochemical will be floating freely in solution, and will not be close enough to the edges of the well to produce signal. Because this is a proximity assay, no wash steps are required (though you can wash if desired).
ScintiPlate microplates are offered with streptavidin-coating, tissue culture-treated (for use in cell-based assays), or uncoated for direct coating procedures. Because they have clear bottom wells, they can be used for bottom-reading or coincidence measurements. Common applications include:
- IRMA sandwich assays - e.g. T4 competitive immunoassay
- RIA competitive assays - e.g. estradiol
- Coated receptor assays - e.g. cloned G-protein or fast ion-receptor assays
- Hybridization assays - e.g. DNA/RNA hybridization or detection of point mutations
- Enzyme activity assays - e.g. kinase with biotinylated substrate
- Cell binding studies - e.g. 2-site antibody to cell-surface receptors
- Binding experiments involving a labelled biotinylated compound - e.g. biotinylated substrate
- Kinetic assays involving a tritiated compound - e.g. an antibody
Cytostar-T plates
Cytostar-T scintillating microplates are sterile, tissue culture treated microplates designed not only for the growth of adherent but also suspension cell cultures. The integral, planar, transparent base of each well is composed of a proprietary homogeneous mixture of scintillants and polystyrene. The transparent nature of the base permits the observation of growth of cells plated in the well. Radioisoes having suitable decay characteristics (3H, 14C, 35S, 45Ca, 125I) brought into proximity with the scintillant contained within the base (via radiochemical uptake or radiochemical interaction with the cells) will have that radioactive decay converted to a light signal. The amount of light generated is proportional to the amount of radioisoe within, or associated with, the cells.
Choosing a scintillating plate
Plate | Use | Measurement | How scintillant is incorporated in plate | Typical application | Notes |
---|---|---|---|---|---|
LumaPlate microplate | Liquid-phase detection | -reading instruments | Bed of solid scintillant on well bottom | Measurement of HPLC fractions | |
Cytostar-T plate | Cell-based assays | - or bottom-reading measurements | Clear scintillating film bottom | Glucose uptake, glucose transport, calcium flux | Clear well bottom allows microscopic visualization of cells |
ScintiPlate microplate | Solid-phase biochemical and cell-based assays | Bottom-reading or coincidence measurements | Scintillant embedded into plastic (sides and bottom of well) | Radioimmunoassays (RIAs), coated receptor assays, cell binding assays, enzymatic assays | Offered with a few pre-coating options, including streptavidin-coated, tissue culture-treated, and uncoated (for coating with a molecule of your choice |
Plate seals
Revvity offers a variety of plate seals. Seal plate seals are applied to the surface of the plate, and are generally used to prevent radioactive contamination or evaporation during assay incubation steps, as well as during plate reading measurements. BackSeals are applied to the bottom of the plate. BackSeal plate seals can be used to seal the bottom of a filter plate prior to the addition scintillation cocktail, preventing leakage. BackSeal plate seals can also be used to change a clear-bottom plate into a white- or black-bottom plate in order to reduce cross-talk during -reading measurements.
Table 4: Plate seal products
Product | Type of seal | Plate format | Number of seals | Catalog number |
---|---|---|---|---|
Seal-A plate seal | Clear adhesive seal | Any | 100 | 6050185 |
Seal-S plate seal | Heat seal for polystyrene | (all) | 100 | 6050192 |
BackSeal plate seal | White adhesive seal | (all) | 55 | 6005199 |
Black adhesive seal | (all) | 55 | 6005189 |
Custom plate services at Revvity
Revvity offers custom microplate services, including bulk ordering, fast and flexible plate barcoding, biological plate coating (including poly-D-lysine, collagen, streptavidin coating, antibody coating, and other coatings on request), custom tissue culture-treatment, custom high protein binding treatment, custom sterilization of microplates, special packaging, and other microplate treatments.
For research use only. Not for use in diagnostic procedures.
The information provided above is solely for informational and research purposes only. Revvity assumes no liability or responsibility for any injuries, losses, or damages resulting from the use or misuse of the provided information, and Revvity assumes no liability for any outcomes resulting from the use or misuse of any recommendations. The information is provided on an "as is" basis without warranties of any kind. Users are responsible for determining the suitability of any recommendations for the user’s particular research. Any recommendations provided by Revvity should not be considered a substitute for a user’s own professional judgment.